Shock Delivery | |
Waveform | |
Shock Series | |
Charge time | -
• Less than 5 seconds to the recommended adult energy level (150 Joules) with a new, fully-charged battery installed -
• Less than 6 seconds to the selected energy level (up to 200 Joules) with a new, fully charged battery installed -
• Less than 15 seconds to the selected energy level while connected to AC power only -
• The device powers on in manual defibrillation mode ready to deliver shock in less than: -
• 15 seconds with a new, fully-charged battery -
• The device powers on in AED mode ready to deliver shock in less than: -
• 24 seconds with a new, fully-charged battery even after 15 discharges of maximum energy |
Sensing for Pads/Paddles | |
Charge times | -
• Less than 5 seconds to the recommended adult energy level (150 Joules) with a new, fully-charged battery installed
• Less than 6 seconds to the selected energy level (up to 200 Joules) with a new, fully charged battery installed
• Less than 15 seconds to the selected energy level while connected to AC power only
• The device powers on in manual defibrillation mode ready to deliver shock in less than:
• 15 seconds with a new, fully-charged battery
• The device powers on in AED mode ready to deliver shock in less than:
• 24 seconds with a new, fully-charged battery even after 15 discharges of maximum energy |
Patient Impedance Range | |